Search
Close this search box.

Products

Features

  1. Good performance of dynamic sealing. Bearing pressure is greater than 37mmHg, clinical routine required 8 ~ 15mmHg.
  2. Resistance of plugging and pulling is very small. Dynamic test data is less than 3.2N/ mm, Similar imported brandsless than 3.2N/mm.
  3. Drag reduction structure of plum membrane flap valve, lower resistance of plugging and pulling.
  4. Puncture cone head has cutting edges with six-sided knife type blade.
  5. Puncture cone universal orientation, without directional insertion.

Technical Specifications

Types and

specifications

Puncture cone outer diameter D1(mm) Casing inner diameter D(mm) Casing length L(mm) Puncture sleeve length L1(mm) Puncture cone length L2(mm)
TC-5/65 5.6 6  

 

65

 

 

105

125
TC-10/65 10.3 10.5      

135

TC-12/65 12.6 13      
TC-15/65 15.4 15.7      
TC-5/100 5.6 6 97  

140

174
TC-10/100 10.3 10.5 98   183
TC-12/100 12.6 13 100   180
TC-15/100 15.6 15.7 111 160 203
Application procedures Apply for laparscopic and surgery procedures, puncture the abdominal tissue of the human body, the establishment ofthe working channel for abdominal surgery