Adult/Pediatric/Neonatal/Medicine/Surgery/Operating Room/ICU/CCU.
| PARAMETER | SPECIFICATIONS | |
| Display | 4’’ real color TFT screen
Resolution: 480*800 One alarm indicator(yellow/red) Standard touch screen |
|
| Power Requirements | AC: 100 ~ 240V, 50Hz/60Hz
DC: Built-in rechargeable battery Battery: 3.7V 2000mAh Fully charged for about 5 hours (single blood oxygen) 5min operating after low battery alarm |
|
| Dimension | Host size: 155*72.5*28.6cm | |
| Storage | Can store 500~1000 sets of historical data | |
| NIBP(Optional) | ||
| Method | Pulse wave oscillometry | |
| Work mode | Manual/ Auto/ STAT | |
| Measure interval of auto mode | 1,2,3,4,5,10,15,30,60,90,120 | |
| Measuring Time of STAT Mode | 5 minutes | |
| PR range | 40 ~ 240bpm | |
| Measure & alarm range: | ||
| Adult | Pediatric | |
| SYS | 40 ~ 270mmHg | 40 ~ 200mmHg |
| DIA | 10 ~ 215mmHg | 10 ~ 150mmHg |
| MEAN | 20 ~ 235mmHg | 20 ~ 165mmHg |
| Static pressure range | 0 ~ 300mmHg | |
| Pressure precision | Max. average error: ±5mmHg
Max. standard deviation: ±8mm Hg |
|
| Over voltage protection | Adult 300mmHg
Pediatric 240mmHg |
|
| Pulse Rate | ||
| Range | 30 ~ 240bpm | |
| Resolution | 1bpm | |
| Accuracy | ±3bpm | |
| SPO2 | ||
| Range | 0 ~ 100% | |
| Resolution | 1% | |
| Precision | 80% ~ 100%: ±2 %
70% ~ 80%: ±3 % 0% ~ 69%: ±no definition given |
|
| ETCO2 (Optional) | ||
| Only side stream | ||
| Warm-up time | when the ambient temperature is 25 ℃, the carbon dioxide curve
(capnogram) can be displayed within 20/15 seconds, and all the specifications can be fulfilled within 2 minutes. |
|
| Measurement range | 0-150mmHg, 0-19.7%, 0-20kPa (at 760mmHg),
atmospheric pressure provided by the host. |
|
| Resolution | 0.1mmHg : 0-69mmHg
0.25mmHg : 70-150mmHg |
|
| Precision | 0-40mmHg : ±2mmHg
41-70mmHg : ±5%(reading) 71-100mmHg : ±8%(reading) 101-150mmHg : ±10%(reading) |
|
| Respiratory rate range | 0- 150 BPM | |
| Respiratory rate accuracy | ±1 BPM | |
| Operating Environment | ||
| Temperature | 0 ~ 40°c | |
| Humidity | ≤85% | |
| Altitude | -500 ~ 4600m | |
| Transport and Storage environment Temperature | -20 ~ 60°c | |
| Humidity | ≤93% | |
| Altitude | -500 ~ 13100m | |
SPO2 sensor, Lithium-ion battery,Power cable, Operator’s manual.