KVO Status
Auto progress KVO mode. Keep the patient’s veins open.
Audible Alarm
Sensitive alarm reminder to ensure injection safety in real time.
One-key Operation
Fulfill conversion between d/min and ml/h.
Pressure Monitoring
Overvoltage protection. The blocking pressure can be detected precisely and sensitively by a pressure sensor installed in the push handle.
Data Storage
Data storage function, convenient for medical staff to review historical alarm information.
Safe and Reliable
System self-test is completed as turning on, which ensures safe work; Real-time bubble monitoring, more secure and reliable.
| PARAMETER | SPECIFICATION |
| Show | 3-inch monochrome LCD display |
| Infusion Speed | 1ml/h – 1200ml/h can be incremented or decreased by 0.1ml/h |
| Infusion Volume Accuracy | ±5% (ordinary infusion set) |
| Infusion Volume Preset | 0.1ml – 9999.9ml can be incremented or decreased by 0.1ml/h |
| Blocking Pressure | High: 110kpa Medium: 80kpa Bottom: 40kpa” |
| KVO | 1ml/h – 5ml/h can be incremented or decreased by 0.1ml/h |
| Voltage | Power supply voltage AC 100~240V AC, power frequency: 50/60Hz
battery input 9.5V-12.6V |
| Power Consumption | Not more than 35VA |
| Built-in Battery Working Time | d.c11.1V no less than 2000mA, no less than 12 hours |
| Dimensions | 90*205*151 |
| Weight | 1.5kg |
| Safety Classification | Class I, type BF general equipment with internal power supply |
| Waterproof Level | IPX1 |
| Working Environment | Ambient temperature: 10~30℃
Relative humidity: 30%~75% Atmospheric pressure: 860hPa–1060hPa The infusion pump should work in an environment where there is no strong shock and vibration, water and other fluids cannot invade the inside of the device, and there is no corrosive gas around. |
| Storage Environment | The packaged infusion pump should be kept in a room with a relative humidity of no more than 93% (no condensation), non-corrosive, good ventilation, and the indoor temperature conditions are: -20℃~+55℃ |