1. Servo Control Mode: air mode
2. Instant Parameter Displaying: set temp, air temp, heat power rate
3. Audible & Visual Alarms: power failure, sensor failure, higher/lower temp, air -temperature override, fan failure
4. Power failure reset memory and self-test function
5. Hood with dual slope: 4hand-ports
Performance | |
Power requirement | ≤400VA |
Temperature control mode | Air servo control |
Air Temp Control Range | 20.0°C-37.0°C |
Temperature fluctuation | ≤0.5°C |
Mattress temperature uniformity | ≤0.8°C |
Temperature rise time | ≤45 min |
Noise level within hood environment | ≤55dB |
Hood | |
Mattress to Hood | 450 mm |
Observation side of hood | Two incline |
Operate port | 4 (2 Iris ports) |
Tubing grommets | 2 |
Front access panel | 826×310 mm |
Mattress Tray Size | 630×360 mm |
Alarms | |
Power failure alarm | |
Fan failure alarm | |
High air temperature alarm: +3.0°C | |
Low air temperature alarm: -3.0°C | |
Over air temperature alarm: ≤38.0°C | |
Operation Condition | |
Environment temperature | +20°C-+30°C |
Environment relative humidity | 30%°C-75% |
Environment air velocity of flow | <0.3m/s |
Other | |
Mattress tray Max. load 5kg | |
Transfusion pole Max. load 2kg |