Products

Features

High accuracy

(1) Tri-angle laser scattering technology

(2) Flow cytometric techniques

(3) 2D scatter plot analysis technology

(4) Cyanide-free HGB measurement technology

Easy operation

(1) One-click return or switch modes

(2) Real-time display of reaction curve

(3) Bidirectional LIS system

Easy Maintenance

(1) Automatic and manual maintenance

(2) Multiple self-checking functions

(3) Modular-design structure

(4) Anti-blocking and one click removal design

Compact Design

(1) Built in fan for automatic cooling

(2) Built-in thermal printer, multiple interfaces

(3) Built-in two kinds of lyse reagents

Technical Specifications

PARAMETER SPECIFICATION
Principle Impedance method for RBC and PLT
Cyanide free colorimetric for HGB
Flow Cytometry+Tri-angle laser scatter + Chemical dye method for WBC 5-part differential analysis and WBC counting
Quality control X-B Floating mean
L-J Quality control
Throughput 60 samples/h
Storage 600,000 test records
Weight 18kg
Throughput 60 samples/hour
Sample volume 20 μL (Whole blood mode)
20 μL (Pre-diluted mode)
Sample type Capillary whole blood, Venous whole blood
Calibration Manual, Calibrator, Fresh blood
Graphic parameters 29 Parameters + 2 Histograms + 2 Scattergrams
Dimensions 325 mm (L) * 475 mm (W) * 465 mm (H)
Weight 28.0 KG
Hematology reagents Diluent B, LH Lyse, DIFF Lyse, Probe cleanser
 

29 Parameters

WBC group (15): WBC, Neu%, Neu#, Lym%, Lym#, Mon%, Mon#, Eos%, Eos#, Bas%, Bas#, *ALY%, *ALY#, *LIC%, *LIC#
RBC group (8): RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV
PLT group (6): PLT, MPV, PDW, PCT, P-LCC, P-LCR

Reliable Results

Parameter Linearity range Repeatability Carryover
WBC (1.0~100)×10⁹/L ≤2.0% (4.0~15.0)×10⁹/L ≤0.5%
RBC (0.00~7.00)×10¹²/L ≤1.5% (3.50~6.00)×10¹²/L ≤0.5%
HGB 20 g/L ~ 200 g/L ≤1.5% (110 g/L~180 g/L) ≤0.6%
PLT (20~1000)×10⁹/L ≤5.0%(100~149)×10⁹/L

≤4.0% (150~500)×10⁹/L

≤1.0%
HCT/MCV ≤1.0% (80fL ~ 110fL) (MCV) ≤0.5% (HCT)