-Manualor Automatic Bolus/Purge/KVO function.
-Vertical & Horizontal mounting clamp.
-DPS dynamic pressure display.
-Anti-bolus system.
-Large LCD display with adjustable brightness.
-2000 events stored can be checked/downloaded.
| PARAMETER | SPECIFICATION |
| Applicable syringes | 5ml,10ml, 20ml, 30ml, 50ml, 60ml of any standard |
| Accuracy | +/- 2% with correct calibration |
| Rate mode | 0.1ml/h-2000.0ml/h (0.1ml/h or 1ml/h step) |
| Time Mode | SOL. VOL: 0.1-9999.9ml (0.1ml step) TIME: 00:01-99:59min (1min step) |
| Dose Mode | Volume: 0.1-999.9ml (0.1ml step) DRUG MASS: 0.1-999.9mg (0.1mg step) DOSE: 0.001-9999 (0.001 step) WEIGHT: 0.1-300kg (0.1kg step) |
| Total Volume Range | 0.1-9999.9ml (0.1ml step) |
| Drug library Mode | (Optional) |
| Sequential Mode | (Optional) |
| Purge Rate | Programmable and adjustable |
| KVO Rate | 0.0~5.0ml/h, Programmable and adjustable |
| DPS Occlusion Pressure | LOW: 60(±15) kPa, MEDIUM: 90(±15) kPa
HIGH: 120(±15) kPa (Adjustable) |
| Alarms | Loose, Occlusion, Empty, Near Empty, Rate Abnormal, Low Battery,
No Battery, Infusion Complete, No Action, Syringe Plunger Disengagement |
| Alarm Volume | Three levels: LOW, MEDIUM or HIGH |
| Bolus Rate | 0.1-2000ml/h, Programmable and adjustable,
Automatic/Manual bolus available |
| Bolus Volume | Range: 1.0~50.0ml, (0.1ml step) |
| Interface | RS-232 |
| Brightness | 1~10 levels can be selected |
| History Records | 2000 events stored can be checked and downloaded |
| Power Source | AC power supply: AC 100 to 240V (±10%), 50/60Hz;
DC Power supply: 12~15V, power>15W; Built-in Lithium Battery: Rechargeable, 11.1V, 2000mAh 12 hours of continuous operation |
| Protection Class | Class I, Type CF, IPX4 |
| Dimension/N.W. | 240mm*150mm*117mm, 2.5kg |
| Packing Size/G.W. | 200mm*260mm*160mm, 3.5kg |
Copyright © 2001-2025 ARI Medical Technology Co., Ltd. - ARI Group. All Rights Reserved.