Products

Feature

1. 15″ TFTTouch Screen High Resolution Display.
2. Resolution ratio: 1024×768.
3. Waveforms: 9 traces on-screen waveforms and maximal upto 13.
4. Changing the color of waveform and parameters.
5. 360-Degree visible indicator with 3- level alarm.
6. Perfusion Index (PI) measurement range: 0.1- 20%.;
7. Integrated handle for easy transfer and storage of accessories.
8. User customized NIBP measuring cycles: up to 5-phase.
9. Versatile clinical calculations for application convenience.;
10. Electrode offset Potential Tolerance: ±800mV.
11. Differential input impedance: ≥10MΩ.;
12. Common Mode Rejection Ratio: Surgery mode, monitor mode, ST mode> 105dB, Diagnostic mode> 90dB
13. Dimension : 357 x 178 x 334 mm
14, Parameters: ECG, RESP, (TEMP), pulse SPO2, PR, NIBP, IBP. ETCO2.

15,Capable of connecting with the Central Monitoring System

Technical Specification

ECG

(1) Sweeping speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s.
(2) Sensitivity selection: x1/8, x1/4, x1/2,x1, x2, x4, Auto
(3) HR measuring range: 15- 350bpm.
(4) HR accuracy: ±1% or ±1bpm,
(5) Pacemaker pulse detection and rejection function.

RESP

(1) Measurement range: Adult 0-120 rmp. Pediatric and neonate: 0 rpm-150rpm.
(2) Measuring accuracy: Adult: 7-120 rpm: ±2 rpm or ±2%,whichever is greater. Pediatric/neonate: 7-150 rpm: ±2 rpm
or ±2%, whichever is greater. 0-6 rpm: not defined

TEMP

(1) Measurement range: 0- 50°C.
(2) Measurement accuracy: ±0.1°C

NIBP

(1) Technique: Oscilometric method.
(2) Typical measurement time: <30 seconds (adult, Child).
(3) NIBP Measuring: SYS, DIA, MAP.
(4) NIBP range:
SYS: 25- 290mmHg (Adult): 25- 240mmHg (Peads) 25- 140mmHg (Neonates).
DIA: 10- 250mmHg (Adult) 10- 200mmHg (Peads) 10- 110mmHg (Neonates).
MAP: 15- 260mmHg(Adult) 15- 215mmHg (Peads) 15- 115mmHg (Neonates).
(5) NIBP measuring accuracy:Mean difference: ±5mmHg. Standard deviation: 8mmHg
(6) Auto intervals: 1- 480min.

SPO2

(1) Technique: Dual-wavelength optical method
(2) Measuring range:0%~100%
(3) Measuring accuracy: Arms is not greater than 2% for SpO2 range 70-100%.
(4) PR measuring range: 30~250bpm
(5) PR measuring accuracy: +2bpm or +1%,
(6) Low perfusion performance: As low as 0.3%.

IBP

(1) Measurement range:-50- 300mmHg.
(2) Measurement positions: ART, RAP, PA, LAP, CVPICP, AUXP1, AUXP2.
(3) Calibration: Zero.
(4) Technique: Strain gauge transducer
(5) Input sensitivity: 5uV/V/mmHg
(6) Measuring accuracy: +2% or +4mmHg, whichever is greater

ETCO2

(1) Technique: Infrared optical method..
(2) Sampling mode: Mainstream.
(3) Range: 0- 150mmHg.
(4) Measurement accuracy: 0- 40mmHg, ±2mmHg, 41- 70mmHg ±5%, 71- 100mmHg ±8%. 101~150mmHg ±10%

Large capacity storage

(1) Up to 2,000 hours of trend data.
(2) Up to 12,000 hrs groups of NIBP records.
(3) 2,000 groups of oxygen- desaturation events.
(4) 2,000 groups of oxygen- desaturation events.
(5) Up to 144 hours of ECG waveform

Power

(1) 220– 240v
(2) Frequency: 50/60 Hz

Standard Packing List

(1) 01 x ECG cable (5 or 3 leads) with leads.
(2) 01 x SPO2 finger sensor for adults/peads.
(3) 01 x NIBP hose.
(4) 01 x NIBP cuff for adult/peads.
(5) 01 x Temperature probe.
(6) 01 x power cord.
(7) 01x operation manual