Products

Features

KVO Status

Auto progress KVO mode. Keep the patient’s veins open.

 

Audible Alarm

Sensitive alarm reminder to ensure injection safety in real time.

 

One-key Operation

Fulfill conversion between d/min and ml/h.

 

Pressure Monitoring

Overvoltage protection. The blocking pressure can be detected precisely and sensitively by a pressure sensor installed in the push handle.

 

Data Storage

Data storage function, convenient for medical staff to review historical alarm information.

 

Safe and Reliable

System self-test is completed as turning on, which ensures safe work; Real-time bubble monitoring, more secure and reliable.

Technical Specifications

PARAMETER SPECIFICATION
Show 3-inch monochrome LCD display
Infusion Speed 1ml/h – 1200ml/h can be incremented or decreased by 0.1ml/h
Infusion Volume Accuracy ±5% (ordinary infusion set)
Infusion Volume Preset 0.1ml – 9999.9ml can be incremented or decreased by 0.1ml/h
Blocking Pressure High: 110kpa Medium: 80kpa Bottom: 40kpa”
KVO 1ml/h – 5ml/h can be incremented or decreased by 0.1ml/h
Voltage Power supply voltage AC 100~240V AC, power frequency: 50/60Hz

battery input 9.5V-12.6V

Power Consumption Not more than 35VA
Built-in Battery Working Time d.c11.1V no less than 2000mA, no less than 12 hours
Dimensions 90*205*151
Weight 1.5kg
Safety Classification Class I, type BF general equipment with internal power supply
Waterproof Level IPX1
Working Environment Ambient temperature: 10~30℃

Relative humidity: 30%~75%

Atmospheric pressure: 860hPa–1060hPa

The infusion pump should work in an environment where there is no strong shock and vibration, water and other fluids cannot invade the inside of the device, and there is no corrosive gas around.

Storage Environment The packaged infusion pump should be kept in a room with a relative humidity of no more than 93% (no condensation), non-corrosive, good ventilation, and the indoor temperature conditions are: -20℃~+55℃