Features

  • Microprocessor controlled, 7-inch LCD touch screen
  • Multifunctional Argon Electrosurgical Generator.
  • Duty Cycle: Intermittent loading and continuous operation (10s/30s)
  • REM system and Power Peak System that provides surgeons with improved performance at lower power settings, minimizing the risk of tissue damage.
  • The newest AIC technology, double-pattern argon current regulation, may respectively output 0.2–3.0 L/min and 1–15 L/min current capacity; the adjustment step length respectively is 0.2 L/min and 1 L/min.
  • We use the SMART INTERRUPT safe system to real-time monitor nozzle pressure, which avoids nozzle jamming.
  • It initiates AUTO-ADJ adaptive pressure adjustment, allowing 0.20–0.50 MPa broad argon indefinite pressure input, avoiding the limitation of the on-the-spot gas source.
  • It offers an effective and economical method for hemostasis and tissue deactivation in open surgeries, such as general surgery, hepatic surgery, thoracic surgery, orthopedic surgery, gynecology and obstetrics surgery, etc.

Technical Specifications

EIGHT WORKING MODES
Function Modes Output Power (load) V-PP CF
Monoploar Cut Pure cut 1-300W @500Ω Load 2.6KV 1.5
Blend 1 1-200W @500Ω Load 4.0KV 1.8
Blend 2 1-150W @500Ω Load 4.2KV 2.1
Monopolar Coagulation Spray 1-80W @500Ω Load 5.3KV 7.4
Forced 1-120W @500Ω Load 5.0KV 3.8
Soft 1-120W @500Ω Load 720V 1.6
Bipolar Standard coag 1-120W @200Ω Load 550V 1.6
Bipolar cut 1-120W @200Ω Load 350V 1.6
INPUT POWER
Input voltage 100-240V~±10%
Mains line frequency range 50-60Hz
Power rating 1100VA
Power consumption ≤1100VA
Fuses (two) 5A / 10A
Working frequency 416KHZ
Monopolar RF leakage current < 150mA
Bipolar RF leakage current < 50mA
ARGON FLOW CONTROL SYSTEM
Flow rate range 0.1~15.0L/min
Regulating step 0.1L/min
Pressure of nozzle 0-0.1Mpa
Power consumption < 100VA
Argon purity ≥99.99%
Argon capacity 4L
Pressure of imputting 0.20-0.50Mpa
AC power AC90-250V,50Hz
OPERTING PARAMETERS
Ambient temperature range 5°C to 40°C
Relative humidity ≤80%
Atmospheric pressure 86kPa to 106kPa
MAIN UNIT DIMENSIONS & WEIGHT
Dimensions L51cm*W42cm*H16cm
Weight 10KGS
STANDARD CONFIGURATION
Main Unit 1 set
Electrosurgical pencil 5 pieces
Electrosurgical pad 10 pieces
Electrosurgical pad cable 1 piece
Monopolar footswitch 1 set
Bipolar footswitch 1 set
Bipolar forceps with cable 1 set
Tip clean sponge 1 piece
Argon pencil 1 piece
Laparoscopic electrode 1 piece
Argon gas probe 1 piece
Electrosurgical snare 1 piece
Argon gas canister 1 bottle
Pressure reducing valve 1 piece
Ventilation connector 1 piece
Trolley 1 set
Power cable 1 piece
Fuse 3 pieces