1. Microprocessor based servo controlled temperature system
2. Control mode: Air mode
3. Humidity is adjustable in two levels
4. Set temperature, air temperature, heating power are displayed separately by LED
5. Self – testing function, various failure alarms by audible and visual
6. Alarm function:
Power failure, Over temperature, Temperature deviation, Fan failure, Temperature sensor failure
7. >37°C temperature set function
8. Triple protection for over temperature with separate cut off device, more safety system
9. The inclination of infant bed is adjustable
10. Single wall hood, 4 operating windows and 2 iris ports, 3 infusion seals
11. Locking device for front door
12. RS232 connector, 1 tray
| Performance | |
| Power supply | AC110/220V, 60/50Hz |
| Power input | 650VA |
| Air temperature control range | 25°C~37°C 37.1°C~38°C |
| Temperature fluctuation | ±0.5°C |
| Uniformity of mattress temperature | ≤0.8°C |
| Warm-up time (from25°C) | <30min |
| Internal noise level | <50dB(A) |
| Infant bed tilt angle | ±10° |
| Mattress size | 65cm(L)*37cm(W) |
| Trough capacity | 1200mL |
| Air filter | 0.5μm |
| Operating Condition | |
| Environment temperature | 20°C~30°C |
| Environment relative humidity | 30%~75% |
| Environment air velocity of flow | <0.3m/s |
| Transport and Storage | |
| Environment temperature | -10°C~+55°C |
| Environment relative humidity | ≤95% |
| Atmospheric Pressure | 500~1060hpa |
| Package | |
| Each unit is packed in one carton | |
| Carton dimension | 128*79*100cm |
| N.W/G.W | 65KG/85KG |
| Standard Configuration | |
| Main body (Including transparent hood,control system,infant bed) | |
| I.V Pole | |
| Mattress | |
| Castors | |
| Tray | |
| Bracket | |
| Optional Functions | |
| AL-3 Upside phototheraphy unit | |
| AL-10 Downside phototherapy unit | |
Copyright © 2001-2025 ARI Medical Technology Co., Ltd. - ARI Group. All Rights Reserved.