| Model | AC9801 |
| Measuring item | K,Na,Cl,Ca,pH |
| Calculating item | nCa, TCa |
| Sample Method | manual |
| Measuring time | 25s |
| Full time | 40s |
| Sample size | 100ul |
| Data storage | 9999 |
| Communication interface | 232 interface |
| Display | 240x64LCD Liquid crystal display,backlight |
| Printer | 58mm thermal printer |
| Air bubble test | Yes |
| Alarm for waste liquid overflow | Yes |
| Alarm for standard liquid level | Yes |
| Standard liquid level indicator lamp | Yes |
| Sample liquid level indicator lamp | Yes |
| Power supply | AC220V±22V or AC110V±11V 50Hz~60Hz 50W |
| Size | 408mmX390mmX230mm |
| Weight | 6KG |
| Measuring method | ISE |
| Working condition | Temperature: 10°C~32°C;Relative humidity ≤85% |
| Measuring item | Blood sample measuring range | Urine sample measuring range | Accuracy | Precision |
| K | 0.50~15.00 mmol/L | 5.0~150.0 mmol/L | ≤±3.0% | CV≤1.5% |
| Na | 30.0~200.0 mmol/L | 10.0~350.0 mmol/L | ≤±3.0% | CV≤1.5% |
| Cl | 30.0~200.0 mmol/L | 10.0~300.0 mmol/L | ≤±3.0% | CV≤1.5% |
| Ca | 0.10~5.00 mmol/L | ──── | ≤±5.0% or ±0.05mmol/L | CV≤1.5% |
| pH | 4.0~9.00 mmol/L | ──── | ≤0.15 | CV≤1.0% |