Display

Display Type: 8 inches High-resolution LCD display.

Sweep Speed: 25mm/sec.

Information: Heart Rate, Lead/Pads, Alarm On/Off, Sp02, AED Functions and Prompts, Alarm Selection and Limits, Delivered Energy.

 

Defibrillator

Waveform: Biphasic

Charge Time: Less than 7 seconds with a new fully charged battery.

Energy Display: Monitor display indicates both selected and delivered energy.

Charge Prompt Type: Voice and visual prompts.

Electrode Impedance Measurement Range: 0-250 ohms.

 

AED Mode

AED Function: Auto analyzes and charges x3 with programmable auto energy level selection, screen prompts, and voice prompts.

Shockable Rhythms: Ventricular fibrillation with amplitude≥200UV, Ventricular tachycardia with rates≥140bpm, and QRS complex wave duration≥140ms.

Charge Control: Control on device front panel, press key on paddle.

Prompts: Voice and visual prompts.

 

Manual Mode

Energy Selection: Selectable at 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 Joules.

Synchronized Mode: Synchronizes defibrillator pulse to patient’s R-wave. “SYNC” message displayed on monitor.

 

ECG Monitoring

Patient Connection: 5-lead ECG cable, or 3-lead ECG cable, paddles.

Lead Selection: Displayed on monitor, paddles, l, II, III, AVR, AVL, AVF, V.

ECG Size: 0.25, 0.5, 1, 1.5, 2, 4 cm/mV display on monitor.

Heart Rate: 20-300BPM.

Heart Rate Alarm: On/Off displayed on monitor, user-selectable.

Smart Alarms: Beeper/Voice prompts indicate shockable rhythm.

 

SpO2 Module

Measurement Range: 30-100%, +2% between 80%-90%, others + 5%

Alarm Range: User set high limit and low limit

Alarm Accuracy: +10% within setting values

Alarm Time Accuracy: Less than 12 sec

 

Recorder

Paper: 50mm thermal.

Speed: 12.5mm/sec, 25mm/sec, 50mm/sec. User-selectable 6-second delay.

Printing Method: High-resolution, thermal array print head.

Print-out Modes: Manual or automatic, user-configurable.

On/Off Control: Front panel and paddle.

Automatic Function: 9-second recording initiated by alarm activation or defibrillator charge or defibrillator discharge.

 

Battery

Type: Rechargeable, Ni-MH battery, 12V.

Operating Time: For a new, fully charged battery: 60 defibrillator discharges at maximum energy, or 5 hours maximum of continuous ECG monitoring.

Additional parameters will affect operating time with different functions.

Option 1

Specification Pacer (Optional)
Type VVI
Pulse Type Rectangular, constant current
Pulse Amplitude 0 to 180 mA±10% or 5 mA (whichever is greater)
Pacing Rate Variable from 30ppm to 180ppm±1.5% (increments or decrements by a value of 2ppm)
Multi-Function Electrode (MFE) Pads Multipurpose defibrillator/pacing electrodes
Pause Pacing pulse frequency reduced by a factor of 4 when activated
Refractory Period NC (VVI demand do not have this specification)
Pulse Width (20+1.5ms)
Output Protection Fully defibrillator protected and isolated

Option 2

Specification EtCO2 (Optional)
Type of sensor By-pass
Technical principle Non-dispersive infrared gas analysis NDIR
Storage condition -40℃ to 70℃, <90℃ RH, non-condensing
Operating conditions 5℃ to 50℃, 10 to 90℃ RH, non-condensing
Ambient pressure 55-115kPa
Power supply 5V±5% (max ripple 200mVp-p)
TDP Typical value 120mA Excursion calibration typical value 280mA
Range 0-19.7% (0-150mmHg, or 0-20Kpa)
Resolution 0.1mmHg
Accuracy 0-40mmHg±2mmHg

41-70mmHg±5% of reading

71-100mmHg±8% of reading

101-150mmHg±10% of reading

Respiratory rate 2-150BPM
Respiratory rate measurement accuracy 1%±1BPM

Option 3

Specification NIBP (Optional)
Measurement unit mmHg/kPa
Measurement Range Adult: 10~200mmHg/kPa

Pediatric: 10~200mmHg/kPa

Neonatal: 10~135mmHg/kPa

Resolution 1mmHg
Accuracy Maximum mean error: 5mmHg

Maximum standard deviation: 8mmHg

Type of sensor By-pass
Technical principle Non-dispersive infrared gas analysis NDIR
Storage condition -40℃ to 70℃, <90% RH, non-condensing
Operating conditions 5℃ to 50℃, 10 to 90% RH, non-condensing
Ambient pressure 55-115kPa
Power supply 5V±5% (max ripple 200mVp-p)
TDP Typical value 120mA Excursion calibration typical value 280mA
Range 0-19.7% (0-150mmHg, or 0-20kPa)
Resolution 0.1mmHg
Accuracy 0-40mmHg±2mmHg

41-70mmHg±5% of reading 71-100mmHg±8% of reading

101-150mmHg±10% of reading

Respiratory rate 2-150BPM
Respiratory rate measurement accuracy 1%±1BPM