| PARAMETER | SPECIFICATION |
| Working Condition | |
| Water input pressure | 0.1Mpa~0.6Mpa, 15P.S.I~60P.S.I |
| Water input temperature | 5℃~30℃ |
| Working environment temperature | 10℃~30℃ at relative humidity ≦70% |
| UF Rate | |
| Flow range | 0ml/h~4000ml/h |
| Resolution ratio | 1ml |
| Precision | ±30ml/h |
| Blood Pump & Substitution Pump | |
| Blood pump flow range | 10ml/min~600ml/min (diameter: 8mm or 6mm) |
| Substitution pump flow range | 10ml/min~300ml/min (diameter 8mm or 6mm) |
| Resolution ratio | 0.1ml |
| Precision | ±10ml or 10% of reading |
| Heparin Pump | |
| Syringe size | 20, 30, 50ml |
| Flow range | 0ml/h~10ml/h |
| Resolution ratio | 0.1ml |
| Precision | ±5% |
| Monitoring System & Alarm Setup | |
| Venous pressure | -180mmHg ~ +600mmHg, ±10mmHg |
| Arterial pressure | -380mmHg ~ +400mmHg, ±10mmHg |
| TMP | -180mmHg ~ +600mmHg, ±20mmHg |
| Dialysate temperature | preset range 34.0℃~39.0℃ |
| Dialysate flow | Less than 800 ml/min (Adjustable) |
| Substitution flow range | 0-28 L/H (on line HDF) |
| Blood leak detection | Photo chromic alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate |
| Bubble detection | Ultrasonic, Alarm when a single air bubble volume is more than 200μl at 200ml/min blood flow |
| Conductivity | Acoustic-optic |
| Disinfection/Sanitize | |
| Hot disinfection | Time: 30 minutes |
| Temperature: about 80℃, at flow rate 500ml/min | |
| Chemical disinfection | Time: 30 minutes |
| Temperature: about 36℃~50℃, at flow rate 500ml/min | |
| Chemical disinfection with heat | Time: 45minutes |
| Temperature: about 36℃~80℃, at flow rate 50ml/min | |
| Rinse | Time: 10minutes |
| Temperature: about 37℃, at flow rate 800ml/min | |
| Storage Environment | Storage temperature should be between 5℃~40℃, at relative humidity ≦80% |
| Power Supply | |
| AC220V, 50Hz/60Hz, 10A | |
| Input power | 1500W |
| Back-up battery | 30 minutes |
| Size &Weight | |
| Size | 380mmx400x1380mm (L*W*H) |
| Net Weight approx. | 88KG |
| Gross Weight approx. | around 100KG |
| Package Size approx | 650×690×1581mm (L x W x H) |
| Function | HDF, on-line BPM, Bi-cart and 2 pcs endotoxin filters |
| Optional Function | Online KT/V, LAN |
Copyright © 2001-2025 ARI Medical Technology Co., Ltd. - ARI Group. All Rights Reserved.