| PARAMETER | SPECIFICATION |
| Main Body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
| Scope of Application | Adult, child |
| Gas Source | O2: 0.28–0.6MP |
| Flow Meter | O2: 0.05–1.0L/min; 1.1–10L/min |
| Flow Rate of Rapid Supply | 25–75L/min |
| Low Oxygen Pressure Alarm | It will be sound alarm when the oxygen pressure < 0.2MPa |
| Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. |
| Respiratory Circuit | Working mode: all-close, semi-close, semi-open
APL: ≤12.5kPa. |
| Respiratory Bellows | Bellows for adults, bellows for children, tidal volume range:0-1500ml |
| VENTILATOR | |
| Display Mode | High-definition 5.7” LCD screen display |
| Ventilation Mode | IPPV, SIPPV, SIMV, MANUAL |
| Ventilation Function | PEEP, SIGH, IRV |
| Ventilation Parameters | Tidal volume 50~1500ml
Rate 2~99bpm SIMV rate 2~20bpm I:E 2:1~1:8 SIGH 1.5 times the inspiratory time/100bpm Inspiratory trigger pressure -10~20cmH2O PEEP 2~20cmH2O Pressure limitation 10~60cmH2O |
| Parameters for Ventilation
Monitoring |
Tidal volume, Minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway-pressure-time waveform, flow rate-time waveform, PEEP, inspiratory trigger pressure, (average pressure, oxygen concentration, battery capacity). |
| SECURITY ALARM SYSTEM | |
| Oxygen Concentration Alarm | Upper limit setting range (21%~99%)
Low limit setting range (18%~80%) |
| Airway Pressure Alarm | Upper limit setting range 55~60cmH2O
Low limit setting range 0~50cmH2O |
| Minute Ventilation Volume Alarm | Upper limit setting range 1~20L/min
Low limit setting range 0~19L/min |
| Tidal volume | Upper limit setting range 100~1500ml
Low limit setting range 0~1400ml |
| Sustained High-pressure
Alarm |
It will give alarm when stress have consistently been higher than
25cmH2O |
| (Suffocation Alarm, Power alarm, Intubation off alarm, battery voltage alarm) | |
| Power | AC 220V, 50Hz, (UPS, Storage battery) |
| (Mechanical Arms) | |
| (End-tidal CO2 Monitoring) | |
| (Specific interface of waste anesthesia gas) | |
| *Note: Items within brackets are optional | |