Products

DEF-7000

Defibrillator Monitor (AED and Manual)

Technical specifications

Display

Display Type: 8 inches High-resolution LCD display.

Sweep Speed: 25mm/sec.

Information: Heart Rate, Lead/Pads, Alarm On/Off, Sp02, AED Functions and Prompts, Alarm Selection and Limits, Delivered Energy.

Defibrillator

Waveform: Biphasic

Charge Time: Less than 7 seconds with a new fully charged battery.

Energy Display: Monitor display indicates both selected and delivered energy.

Charge Prompt Type: Voice and visual prompts.

Electrode Impedance Measurement Range: 0-250 ohms.

AED Mode

AED Function: Auto analyzes and charges x3 with programmable auto energy level selection, screen prompts, and voice prompts.

Shockable Rhythms: Ventricular fibrillation with amplitude200UV, Ventricular tachycardia with rates140bpm, and QRS complex wave duration140ms.

Charge Control: Control on device front panel, press key on paddle.

Prompts: Voice and visual prompts.

Manual Mode

Energy Selection: Selectable at 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 Joules.

Synchronized Mode: Synchronizes defibrillator pulse to patient's R-wave. "SYNC" message displayed on monitor.

ECG Monitoring

Patient Connection: 5-lead ECG cable, or 3-lead ECG cable, paddles.

Lead Selection: Displayed on monitor, paddles, l, II, III, AVR, AVL, AVF, V.

ECG Size: 0.25, 0.5, 1, 1.5, 2, 4 cm/mV display on monitor.

Heart Rate: 20-300BPM.

Heart Rate Alarm: On/Off displayed on monitor, user-selectable.

Smart Alarms: Beeper/Voice prompts indicate shockable rhythm.

SpO2 Module

Measurement Range: 30-100%, +2% between 80%-90%, others + 5%

Alarm Range: User set high limit and low limit

Alarm Accuracy: +10% within setting values

Alarm Time Accuracy: Less than 12 sec

Recorder

Paper: 50mm thermal.

Speed: 12.5mm/sec, 25mm/sec, 50mm/sec. User-selectable 6-second delay.

Printing Method: High-resolution, thermal array print head.

Print-out Modes: Manual or automatic, user-configurable.

On/Off Control: Front panel and paddle.

Automatic Function: 9-second recording initiated by alarm activation or defibrillator charge or defibrillator discharge.

Battery

Type: Rechargeable, Ni-MH battery, 12V.

Operating Time: For a new, fully charged battery: 60 defibrillator discharges at maximum energy, or 3 hours minimum of continuous ECG monitoring.

Additional parameters will affect operating time with different functions.

Option 1

Specification

Pacer (Optional)

Type

VVI

Pulse Type

Rectangular, constant current

Pulse Amplitude

0 to 180 mA±10% or 5 mA (whichever is greater)

Pacing Rate

Variable from 30ppm to 180ppm±1.5% (increments or decrements by a value of 2ppm)

Multi-Function Electrode (MFE) Pads

Multipurpose defibrillator/pacing electrodes

Pause

Pacing pulse frequency reduced by a factor of 4 when activated

Refractory Period

NC (VVI demand do not have this specification)

Pulse Width

(20+1.5ms)

Output Protection

Fully defibrillator protected and isolated

Option 2

Specification

EtCO2 (Optional)

Type of sensor

By-pass

Technical principle

Non-dispersive infrared gas analysis NDIR

Storage condition

-40 to 70, <90 RH, non-condensing

Operating conditions

5 to 50, 10 to 90 RH, non-condensing

Ambient pressure

55-115kPa

Power supply

5V±5% (max ripple 200mVp-p)

TDP

Typical value 120mA Excursion calibration typical value 280mA

Range

0-19.7% (0-150mmHg, or 0-20Kpa)

Resolution

0.1mmHg

Accuracy

0-40mmHg±2mmHg

41-70mmHg±5% of reading

71-100mmHg±8% of reading

101-150mmHg±10% of reading

Respiratory rate

2-150BPM

Respiratory rate measurement accuracy

1%±1BPM

Option 3

Specification

NIBP (Optional)

Measurement unit

mmHg/kPa

Measurement Range

Adult: 10~200mmHg/kPa

Pediatric: 10~200mmHg/kPa

Neonatal: 10~135mmHg/kPa

Resolution

1mmHg

Accuracy

Maximum mean error: 5mmHg

Maximum standard deviation: 8mmHg

Type of sensor

By-pass

Technical principle

Non-dispersive infrared gas analysis NDIR

Storage condition

-40 to 70, <90% RH, non-condensing

Operating conditions

5 to 50, 10 to 90% RH, non-condensing

Ambient pressure

55-115kPa

Power supply

5V±5% (max ripple 200mVp-p)

TDP

Typical value 120mA Excursion calibration typical value 280mA

Range

0-19.7% (0-150mmHg, or 0-20kPa)

Resolution

0.1mmHg

Accuracy

0-40mmHg±2mmHg

41-70mmHg±5% of reading 71-100mmHg±8% of reading

101-150mmHg±10% of reading

Respiratory rate

2-150BPM

Respiratory rate measurement accuracy

1%±1BPM

 